User manual THOMPSON SURGICAL PANNICULECTOMY RETRACTOR SYSTEM

DON'T FORGET : ALWAYS READ THE USER GUIDE BEFORE BUYING !!!

If this document matches the user guide, instructions manual or user manual, feature sets, schematics you are looking for, download it now. Diplodocs provides you a fast and easy access to the user manual THOMPSON SURGICAL PANNICULECTOMY RETRACTOR SYSTEM. We hope that this THOMPSON SURGICAL PANNICULECTOMY RETRACTOR SYSTEM user guide will be useful to you.


THOMPSON SURGICAL PANNICULECTOMY RETRACTOR SYSTEM : Download the complete user guide (1832 Ko)

Manual abstract: user guide THOMPSON SURGICAL PANNICULECTOMY RETRACTOR SYSTEM

Detailed instructions for use are in the User's Guide.

[. . . ] Thompson User Manual Panniculectomy Retractor System jckdpanum0307 1 Opposite the surgeon, at the patient's nipple line, secure the Elite II Rail Clamp (#41902AC) over the sterile drape to the OR table. 2 Insert the Collapsible Grooved Rod (#44734) into the Elite II Rail Clamp's cam joint with the grooved portion in the horizontal position. [. . . ] Lock into place by actuating the handle and pressing it tight. 5 Attach Perforating Towel Clips to pannus as needed and loop the finger rings over the Towel Clip Crank hooks. Retract as needed by turning the Towel Clip Crank's knobs. jckdpanum0307 C A R E & MAINTENANCE CLEAN ALL COMPONENTS. Ultrasonic cleaning is recommended; however, hand washing with a brush is acceptable. Be certain that all Cam Joints are in FULL OPEN POSITION during autoclaving. STERILIZE: GRAVITY STEAM (Wrapped) 30 minutes 121°C (250°F) @ 1. 1kg/cm2 HIGH SPEED GRAVITY (Flash) 5 minutes 132°C (270°F) @30psi or 132° @ 2kg/cm2 PREVACUUM STEAM (Wrapped) 10 minutes @ 270°F or 132°C These sterilization settings are recommendations. Consult your autoclave manual for specific instructions, conditions and exposure periods. FOLLOWING STERILIZATION, ALLOW YOUR INSTRUMENTS TO COOL BEFORE HANDLING. [. . . ] Elite II Rail Clamp with Cam Joint 41902AC Towel Clip Cranks 35028 10170 East Cherry Bend Road Traverse City, Michigan 49684 (800) 227-7543 | Fax (231) 922-0174 For Customer Service call (800) 227-7543 or (231) 922-0177 Or Visit us on the World Wide Web at Straight Arm with 1 Cam Joint 44120C www. thompsonsurgical. com 90° Collapsible Rod with Grooves 44734 jckdpanum0307 [. . . ]

DISCLAIMER TO DOWNLOAD THE USER GUIDE THOMPSON SURGICAL PANNICULECTOMY RETRACTOR SYSTEM




Click on "Download the user Manual" at the end of this Contract if you accept its terms, the downloading of the manual THOMPSON SURGICAL PANNICULECTOMY RETRACTOR SYSTEM will begin.

 

Copyright © 2015 - manualRetreiver - All Rights Reserved.
Designated trademarks and brands are the property of their respective owners.