If this document matches the user guide, instructions manual or user manual, feature sets, schematics you are looking for, download it now. Diplodocs provides you a fast and easy access to the user manual HANNA INSTRUMENTS HI 931001. We hope that this HANNA INSTRUMENTS HI 931001 user guide will be useful to you.
Manual abstract: user guide HANNA INSTRUMENTS HI 931001
Detailed instructions for use are in the User's Guide.
[. . . ] These instruments are in compliance with the directives EN 50081-1 and EN 50082-1.
PRELIMINARY EXAMINATION Remove the instrument from the packing material and examine it carefully to make sure that no damage has occurred during shipment. Each pH/mV simulator is supplied complete with: · HI 7858/1 BNC/BNC Coaxial Cable; Note: Save all packing material until the instrument has been observed to function correctly because all defective items must be returned to the Dealer in their original packing.
TABLE OF CONTENTS Preliminary Examination . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 GENERAL DESCRIPTION HI 8427 and HI 931001 are portable pH/mV simulators designed specifically for calibrating pH and mV (ORP) meters. [. . . ] pH 4 at 30°C).
pH7
HI 931001: · Connect the simulator to the pH meter using the HI 7858/1 cable provided. · Manually set the temperature compensation of your meter to 25°C.
· Switch the simulator on and select the pH mode by pressing the RANGE key.
pH
pH
HI 8424
microcomputer pHmeter
ON OFF
HI 931001
pH/mV Calibrator
SETTING
ON OFF
RANGE
pH
HI 8427 pH/mV CALIBRATOR
·
0
10 20 30
Make sure that the two additional decimal points are not displayed. If they are displayed, change the battery of the simulator before performing the calibration. Turn the Coarse and Fine knob to set any desired pH value and make sure that the pH meter you are checking reflects the correct result (e. g. 7. 01 pH).
·
50 40 °C
·
·
If your pH meter has a temperature probe connected to it, set the °C knob to the measured temperature displayed by your pH meter. If your meter is manually compensated for temperature, adjust the temperature settings for both the pH meter and the simulator to the same temperature. Verify that the pH meter displays the correct pH value compensated at the
6
·
If not, adjust the pH meter until the correct reading is displayed (refer to the pH meter instruction manual).
pH
pH
HI 8424
microcomputer pHmeter
ON OFF
HI 931001
pH/mV Calibrator
SETTING
7
mV (ORP) CALIBRATION
·
HI 8427: · Connect the simulator to the mV (ORP) meter using the HI 7858/1 cable provided.
·
·
pH
HI 8427 pH/mV CALIBRATOR
Switch the mV meter on and select the desired mV value on the simulator. If it is lit, change the bat-350 tery of the simulator bemV fore performing the calibration. · If the meter is not displaying the correct mV value, adjust the meter until it displays the correct mV (refer to the mV meter instructions manual). Note: The °C knob is not effective during mV calibration.
Make sure that the the two additional decimal points are not displayed. If they are displayed, change the battery of the simulator before performing the calibration. Turn the Coarse or Fine knob to any desired value and make sure that the mV meter you are checking reflects the correct result.
·
If not, adjust the mV meter until the correct reading is displayed (refer to the mV meter instruction manual).
mV
mV
HI 8424
microcomputer pHmeter
ON OFF
HI 931001
pH/mV Calibrator
SETTING
Note: Temperature compensation does not apply here.
HI 931001: · Connect the simulator to the mV (ORP) meter using the HI 7858/1 cable provided.
pH
pH
HI 8424
microcomputer pHmeter
ON OFF
HI 931001
pH/mV Calibrator
SETTING
ON
· Switch the simulator on and select the mV mode by pressing the RANGE key.
OFF
RANGE
8
9
CABLE HIGH IMPEDANCE TEST
pH VALUES AT VARIOUS TEMPERATURE
HI 8427 only:
Temperature has an effect on pH. The calibration buffer solutions are effected by temperature changes to a lesser degree than normal solutions (see table below). Note: With HI 8427 your pH meter will always show the pH buffer solution at the standard temperature of 25ºC.
·
Connect the simulator to the cable (e. g. HI 7855).
·
Press the HIGH IMPEDANCE key and observe the LEAKAGE LED while you are turning the simulator rotary knob to any pH/mV value.
HIGH IMPEDANCE
·
If the LED does not light, the cable and connector going to the meter are in perfect condition. If the LED is lit, it means there is a leakage problem with the cable and connector and you should check/replace it.
LEAKAGE
TEMP °C °F
0 5 10 15 20 25 30 35 40 45 50 55 60 65 70 32 41 50 59 68 77 86 95 104 113 122 131 140 149 158
4. 01
4. 01 4. 00 4. 00 4. 00 4. 00 4. 01 4. 02 4. 03 4. 04 4. 05 4. 06 4. 07 4. 09 4. 11 4. 12
pH VALUES 6. 86 7. 01 9. 18
6. 98 6. 95 6. 92 6. 90 6. 88 6. 86 6. 85 6. 84 6. 84 6. 83 6. 83 6. 84 6. 84 6. 85 6. 85 7. 13 7. 10 7. 07 7. 04 7. 03 7. 01 7. 00 6. 99 6. 98 6. 98 6. 98 6. 98 6. 98 6. 99 6. 99 9. 46 9. 39 9. 33 9. 27 9. 22 9. 18 9. 14 9. 10 9. 07 9. 04 9. 01 8. 99 8. 97 8. 95 8. 93
10. 01
10. 32 10. 24 10. 18 10. 12 10. 06 10. 01 9. 96 9. 92 9. 88 9. 85 9. 82 9. 79 9. 77 9. 76 9. 75
·
LEAKAGE
10
11
BATTERY REPLACEMENT HI 710001 The instruments use a 9 volt battery that lasts for approximately 100 hours or 500 hours of continuous use with HI 8427 and HI 931001 respectively. In the event of dead battery the LOW BATTERY LED of HI 8427 will be lit or two additional decimal points will be displayed on HI 931001.
LOW BATTERY LEAKAGE
ACCESSORIES Soft carrying case for HI 931001, dimensions 230 x 100 x 50 mm Extension cable with 3 mm diameter (0. 12"), 1m (3. 3') long Extension cable with 3 mm diameter (0. 12"), 3m (9. 9') long Extension cable with 3 mm diameter (0. 12"), 5m (16. 5') long
HI 7858/1 HI 7858/3 HI 7858/5
pH
HI 7858/10 Extension cable with 3 mm diameter (0. 12"), 10m (33') long HI 7858/15 Extension cable with 3 mm diameter (0. 12"), 15m (49. 5') long
HI 7858 SERIES CABLE CONNECTORS BNC AND 3. 0 mm (0. 12") CABLE WITH BNC
Battery replacement must only take place in a non hazardous area using the battery types specified in this instruction manual (see page 11). [. . . ] To validate your warranty, fill out and return the enclosed warranty card within 14 days from the date of purchase.
CE DECLARATION OF CONFORMITY
Recommendations for Users
Before using these products, make sure that they are entirely suitable for the environment in which they are used. Operation of these instruments in residential area could cause unacceptable interferences to radio and TV equipments, requiring the operator to take all necessary steps to correct interferences. Any variation introduced by the user to the supplied equipment may degrade the instruments' EMC performance. [. . . ]