Instruction Manual
China:
Tel. (10) 88570068 · Fax (10) 88570060
HI 8424NEW
Portable pH/mV/°C Meter with Automatic Calibration and Battery Recharging System
Egypt:
Tel. 886. 2. 2739. 3014 · Fax 886. 2. 2739. 2983
MAN8424W R1
Thailand:
Tel. (662) 619. 0708. 11 · Fax (662) 619. 0061
United Kingdom:
Tel. [. . . ] The displayed pH reading will be temperature compensated at the set value (in this case at 25 °C).
FC 200B / FC 200S PVDF body, single junction, conic, Viscolene, non-refillable, combination pH electrode. Use: meat & cheese.
6 mm 0. 25"
FC 200
75 mm 2. 95"
FC 210B Glass body, double junction, conic, Viscolene, non-refillable combination pH electrode. Use: milk, yogurt.
12 mm 0. 5"
ORP MEASUREMENTS Oxidation Reduction Potential (ORP) measurements provide a quantification of the oxidizing or reducing power of the sample tested. · To enter the "mV" mode turn the instrument ON and press the RANGE key until the display changes to mV. · Submerge the ORP electrode tip (at least 4 cm / 1½") into the sample to be tested and allow time for the reading to stabilize (hourglass symbol turns off). · Measurements within the ±699. 9 mV range are displayed with 0. 1 mV resolution, while outside this range the resolution is 1 mV.
FC 210 120 mm 4. 7"
22
7
Notes: · To perform correct ORP measurements, the surface of the ORP electrode must be clean and smooth. · When not in use, the tip of the electrode should be kept moist (use HI 70300 storage solution) and safe from any mechanical stress which might cause damage to the glass/platinum junction. TEMPERATURE MEASUREMENTS · Turn the instrument ON and press the RANGE key to select the temperature mode. · Dip the temperature probe into the sample, allow the reading to stabilize (hourglass symbol turns off) and read the temperature value. · Temperature measurements can be displayed in °C or °F units (see "Menu selection" for details).
HI 1083B Glass body, micro, Viscolene, non-refillable, combination pH electrode. Use: biotechnology, micro titration.
12 mm 0. 5" 5 mm 0. 2" 3 mm 0. 12" 3. 0 mm DIA 0. 12"
HI 1083
120 mm 4. 7"
HI 1131B / HI 1111S Glass body, single junction, refillable, combination pH electrode. Use: general purpose.
9. 5mm DIA 0. 37" 12 mm 0. 5"
HI 1131 120 mm 4. 7"
"S" VERSION
Notes: · A blinking full scale value means that the reading is out of range. · To freeze a reading on display while in measurement mode, press the HOLD key. The pH, mV and temperature values are held, and the RANGE key can be used to view the values. · To save battery life, the meter is provided with an auto-off feature, which turns the instrument off after 20 minutes of non-use. This feature can be disabled by the user (see "Menu selection" for details).
HI 1230B / HI 1210S Plastic body, double junction, gel-filled, combination pH electrode. Use: general purpose.
12 mm 0. 5"
HI 1230
120 mm 4. 7"
"S" VERSION
HI 1330B / HI 1310S Glass body, semimicro, single junction, refillable, combination pH electrode. Use: laboratory.
5mm DIA 0. 2" 5mm 0. 2"
HI 1330 120 mm 4. 7"
"S" VERSION
HI 1331B / HI 1311S Glass body, semimicro, single junction, refillable, combination pH electrode. Use: flasks.
8 mm 0. 3" 7. 5mm DIA 0. 29"
HI 1331
210 mm 8. 25"
"S" VERSION
8
21
HI 7082 HI 8082 HI 8093
3. 5 M KCl electrolyte solution, 4 x 30 mL bottle, for double junction electrodes 3. 5 M KCl electrolyte solution, 4 x 30 mL FDA bottle, for double junction electrodes 1 M KCl+AgCl electrolyte solution, 4 x 30 mL FDA bottle, for double junction electrodes
p H CALIBRATION
For better accuracy, frequent calibration of the instrument is recommended. The instrument should be recalibrated for pH: a) Whenever the pH electrode or temperature probe is replaced b) At least once a week c) After testing aggressive chemicals d) When extreme accuracy is required PREPARATION Pour small quantities of pH 7. 01 (HI 7007) and pH 4. 01 (HI 7004) or pH 10. 01 (HI 7010) buffer solutions into two clean beakers. For accurate calibration use two beakers for each buffer solution, one for rinsing the electrode tip, and one for calibration. For measurements in acidic samples, it is recommended to calibrate the meter by using pH 7. 01 (HI 7007) and pH 4. 01 (HI 7004) buffers, while for alkaline measurements use pH 7. 01 (HI 7007) and pH 10. 01 (HI 7010) buffers. [. . . ] The LCD shows in three subsequent screens "Auto", "OFF" and "20" to indicate that the 20 minutes selection is active, or "Auto", "OFF" and "no" if the feature is disabled.
ELECTRODE CONDITIONING & MAINTENANCE
The user can toggle the selection by pressing CAL/SELECT or move to the next step with CFM/SETUP. The following selection is the acoustic signal, which is displayed on two subsequent screens: "bEEP", "OFF" when the feature is disabled, and "bEEP", "On" when the feature is enabled.
Press CAL/SELECT to toggle the selection and CFM/SETUP to move to next step. At this point it is possible to set the temperature unit, by selecting "Unit", "°C" or "Unit", "°F".
Only for refillable electrodes; must be open while measuring. This is normal with electrodes and they will disappear when rinsed with water. [. . . ]