User manual HANNA INSTRUMENTS HI 755

DON'T FORGET : ALWAYS READ THE USER GUIDE BEFORE BUYING !!!

If this document matches the user guide, instructions manual or user manual, feature sets, schematics you are looking for, download it now. Diplodocs provides you a fast and easy access to the user manual HANNA INSTRUMENTS HI 755. We hope that this HANNA INSTRUMENTS HI 755 user guide will be useful to you.


HANNA INSTRUMENTS HI 755 : Download the complete user guide (646 Ko)

Manual abstract: user guide HANNA INSTRUMENTS HI 755

Detailed instructions for use are in the User's Guide.

[. . . ] Each HI 755 meter is supplied complete with: · Two Sample Cuvettes and Caps · Six powder reagents for Alkalinity · 1 x 1. 5V AAA Battery · Instruction Manual 1. Button Errors and warnings: Light High: There is too much light to perform a measurement. Under range: A blinking "0" indicates that the sample absorbs less light than the zero reference. [. . . ] Once this indication is displayed, normal operation of the instrument will be interrupted. Change the battery and restart the meter. Measurement procedure: · Turn the meter on by pressing the button. After all the segments are displayed, "C. 1", "Add" appears with "Press" blinking, the meter is ready. When the display shows "Add", "C. 2" with "Press" blinking the meter is zeroed. Tips for an accurate measurement · It is important that the sample does not contain any debris. · Whenever the cuvette is placed into the measurement cell, it must be dry outside, and completely free of fingerprints, oil or dirt. Wipe it throughly with HI 731318 or a lint-free cloth prior to insertion. · Shaking the cuvette can generate bubbles, causing higher readings. To obtain accurate measurements, remove such bubbles by swirling or by gently tapping the cuvette. · Do not let the reacted sample stand for too long after reagent is added, or accuracy will be lost. · After the reading it is important to discard the sample immediately , otherwise the glass might become permanently stained. Accessories: REAGENT SETS HI 755-25 OTHER ACCESSORIES HI 740028 HI 731318 HI 731321 HI 731225 HI 93703-50 Reagents for 25 Alkalinity tests 1. 5V AAA batteries (4 pcs) Tissue for wiping cuvettes (4 pcs) Glass cuvettes (4 pcs) Cuvette cap for checker HD (5 pcs) Cuvettes cleaning solution (230 mL). · Recommendations for Users Before using these products, make sure that they are entirely suitable for your specific application and for the environment in where they are used. Operation of these instruments may cause unacceptable interferences to other electronic equipments, thus requiring the operator to take all necessary steps to eliminate interferences. Any variation introduced by the user to the supplied equipment may degrade the instrument's EMC performance. To avoid damage or burns, do not put the instrument in the microwave oven. For your safety do not use or store the instrument in hazardous environments. Battery management To save the battery, the instrument shuts down after 10 minutes of non-use. [. . . ] Wait until the message "C. 2" appears with "Press" blinking, then invert gently again 3 times the cuvette, avoid bubble development inside. The instrument directly displays the concentration of alkalinity in ppm of CaCO 3. Alkalinity conversion: 1 ppm CaCO 3/L = 0. 02 meq/L = 0. 056 dkH The meter automatically turns off after 10 minutes. Hanna Instruments reserves the right to modify the design, construction and appearance of its products without advance notice. For additional information, contact your dealer or the nearest Hanna Customer Service Center. [. . . ]

DISCLAIMER TO DOWNLOAD THE USER GUIDE HANNA INSTRUMENTS HI 755




Click on "Download the user Manual" at the end of this Contract if you accept its terms, the downloading of the manual HANNA INSTRUMENTS HI 755 will begin.

 

Copyright © 2015 - manualRetreiver - All Rights Reserved.
Designated trademarks and brands are the property of their respective owners.