User manual HANNA INSTRUMENTS HI 110
DON'T FORGET : ALWAYS READ THE USER GUIDE BEFORE BUYING !!!
If this document matches the user guide, instructions manual or user manual, feature sets, schematics you are looking for, download it now. Diplodocs provides you a fast and easy access to the user manual HANNA INSTRUMENTS HI 110. We hope that this HANNA INSTRUMENTS HI 110 user guide will be useful to you.
Manual abstract: user guide HANNA INSTRUMENTS HI 110
Detailed instructions for use are in the User's Guide.
[. . . ] Instruction Manual
HI 110 HI 112
HI 111 HI 113
pH/mV/°C Bench Meters with Custom Buffers
pH
CAL
CF GLPM
MR
Func tions
GLP MEM
Show
C C MEM MR
RANG SET
Store
GLP ing
Memo ry Re call Selec Selec t pH/O RP Set Pa t Setup Ite m ramete rs
E
Read
Data
pH Ca
CAL
libratio n
/
CFM
Selec Confi
HI pH / 113 ORP
Enter /Exit Calib ration t pH Buffe r rm Ca librat ion Po int
RANG
E
Met er
SET
w w w. c o m
1
Dear Customer, Thank you for choosing a Hanna Instruments product. Please read this instruction manual carefully before using the instruments. This manual will provide you with the necessary information for correct use of the instruments, as well as a precise idea of their versatility. [. . . ] The instruments will display the time (hh:mm) of the last calibration.
These instruments allow the user to set the number of days before the next required calibration. Note: If the instruments were not calibrated, the CAL INTV tags will be displayed even if the feature is disabled in SETUP menu. LAST CALIBRATION DATA
Press the ARROW keys to view the next logged calibration parameter (pressing the ºC key): The date (yyyy. mm. dd).
CAL DATE
The pH calibration offset.
CAL mV
The pH calibration slope (the GLP slope is the average of the calibration slopes; the percentage is referred to the ideal value of 59. 16 mV/pH).
CAL %
15
The pH calibration buffers in calibrating order. The first pH calibration buffer:
CAL pH BUFFER pH
BUF
The second pH calibration buffer:
CAL pH BUFFER pH
BUF
The third pH calibration buffer:
CAL pH BUFFER pH
BUF
Note:
For HI 112 and HI 113, the fourth and the fifth calibration buffers are also displayed in GLP. The OLd message displayed beside the pH value means that this buffer was not used during last calibration. Press and hold down SET if you want to see calibration date (or time, if old calibration was made in the same day with current calibration). For each custom buffer used in calibration, the BUFFER pH tag will blink. If no bUF message appears on the LCD, the instrument informs you that calibration was performed in less than three (HI 110, HI 111) or five points (HI 112, HI 113).
CAL
BUF
pH resolution during calibration.
CAL
Calibration Alarm Time-Out status: if disabled.
16
or the number of days until the calibration alarm will be displayed.
or if expired (7 days ago).
CAL INTV
The instrument ID.
Note:
Press GLP at any moment and the instruments will return to measurement mode. If calibration has not been performed, the instruments display no CAL message blinking.
CAL
Setup mode allows viewing and modifying the following parameters: Calibration Alarm Time-Out Custom buffer 1 Custom buffer 2 One-point calibration behaviour Instrument ID (HI 112 & HI 113) Current Time Current Date Beep Status Baud Rate (serial communication, HI 112 & HI 113) Command prefix (serial communication, HI 112 & HI 113) To enter SETUP mode press SET while the instrument is in measurement mode. 17
SETUP
CALIBRATION ALARM TIME-OUT SET
Press CAL when the calibration time-out is displayed. Calibration time-out (OFF or 1 to 7 days) will start blinking.
CAL INTV
Press the ARROWkeys to change the calibration time-out value. FIRST CUSTOM BUFFER SET
Press CAL when custom buffer 1 is displayed. The custom buffer (disabled no cb1 or 0 to 16 pH) will start blinking.
CAL BUFFER pH
Press CAL when custom buffer 2 is displayed. The custom buffer (disabled no cb2 or 0 to 16 pH) will start blinking.
CAL BUFFER pH
Press the ARROW keys to change the custom buffer value. SECOND CUSTOM BUFFER SET
Press the ARROW keys to change the custom buffer value. ONE-POINT CALIBRATION BEHAVIOUR
Press CAL when 1 Pnt message is displayed on the secondary LCD. One of the two options (Pnt or OFFS) will start blinking (see pH CALIBRATION PROCEDURE for details, page 11).
CAL
Press the ARROW keys to toggle between Pnt and OFFS options. 18
INSTRUMENT ID PARAMETER SET (HI 112, HI 113 only)
Press CAL when the instrument ID is displayed. The instrument ID (0000 to 9999) will start blinking.
Press the ARROWkeys to change the instrument ID value. The hour will start blinking.
TIME
Press the ARROWkeys to change the hour. The minutes will start blinking.
TIME
Press the ARROW keys to change the minutes. The year will start blinking.
DATE
Press the ARROWkeys to change the year. [. . . ] Use: biotechnology, micro titration.
HI 1050
12 mm 0. 5" 5 mm 0. 2" 3 mm 0. 12" 3. 0 mm DIA 0. 12"
HI 1083
120 mm 4. 7"
HI 1131B Glass-body, single junction, refillable, combination pH electrode. Use: general purpose.
9. 5mm DIA 0. 37" 12 mm 0. 5"
HI 1131 120 mm 4. 7"
"S" VERSION
34
HI 1111
HI 1330B Glass-body, semimicro, single junction, refillable, combination pH electrode. Use: laboratory, vials.
5mm DIA 0. 2" 5mm 0. 2"
HI 1330 120 mm 4. 7"
HI 1331B Glass-body, semimicro, single junction, refillable, combination pH electrode. Use: flasks.
HI 1310
"S" VERSION
8 mm 0. 3"
7. 5mm DIA 0. 29"
HI 1331
210 mm 8. 25"
HI 1230B Plastic-body (Ultem®), double junction, gel-filled, combination pH electrode. [. . . ]