User manual HANNA INSTRUMENTS HI 983315-0

DON'T FORGET : ALWAYS READ THE USER GUIDE BEFORE BUYING !!!

If this document matches the user guide, instructions manual or user manual, feature sets, schematics you are looking for, download it now. Diplodocs provides you a fast and easy access to the user manual HANNA INSTRUMENTS HI 983315-0. We hope that this HANNA INSTRUMENTS HI 983315-0 user guide will be useful to you.


HANNA INSTRUMENTS HI 983315-0 : Download the complete user guide (358 Ko)

Manual abstract: user guide HANNA INSTRUMENTS HI 983315-0

Detailed instructions for use are in the User's Guide.

[. . . ] These instruments are in compliance with the directives. SPECIFICATIONS Range 0. 0 to 199. 9 ppm Resolution 0. 1 ppm Accuracy (@ 20°C) ±2% f. s. Setpoint Adjustable through multiturn trimmer Alarm Condition LED ON and alarm contact closed when measured TDS value is higher than setpoint Alarm Output 2-contact relay, no fuse protected. Compensation Automatic from 5 to 50°C (41 to 122°F) with ß=2% TDS Conversion Factor 0. 5 Calibration Manual, at one point through trimmer Calibration Solution HI 7033 (84 µS/cm) Power supply: External HI983315-0 12 VDC HI98331500 12VDC isolated inside with DC/DC converter HI983315-1 115/230 VAC; 50/60Hz Dimensions 79 x 49 x 95 mm (3. 1 x 1. 9 x 3. 7") Weight: HI983315-0, HI98331500 HI983315-1 140 g (4. 96 oz) 250 g (8. 85 oz) CE DECLARATION OF CONFORMITY PRELIMINARY EXAMINATION Remove the instrument from the packing material and examine it carefully. [. . . ] Hanna Instruments reserves the right to modify the design, construction and appearance of its products without advance notice. GENERAL DESCRIPTION HI983315-0, HI983315-1 and HI98331500 are conductivity indicators and controllers with a relay output designed for simplicity of use in a wide range of industrial applications. HI98331500 is provided with a DC/DC converter which isolates the measure system from the external power supply. They are panel-mounted with membrane keypads on the front panel and an easy-to-read LCD display. Measurements are highly accurate and the meters can be calibrated at one point. Power supply, wiring and selection are made via the plugin terminal blocks on the rear panel. LED indicators on the front panel identify whether the controller is in set or reading mode; alarm led lights up when measurements are higher than selected setpoint value. Recommendations for Users Before using these products, make sure that they are entirely suitable for the environment in which they are used. Operation of these instruments in residential areas could cause unacceptable interferences to radio and TV equipment. The metal band at the end of the probe is sensitive to electrostatic discharges. During operation, ESD wrist straps should be worn to avoid possible damage to the probe by electrostatic discharges. Any variation introduced by the user to the supplied equipment may degrade the instrument's EMC performance. To avoid electrical shock, do not use these instruments when voltage at the measurement surface exceeds 24 VAC or 60 VDC. To avoid damage or burns, do not perform any measurement in microwave ovens. ISTR983315R5 12/2001 http://www. hannainst. com FUNCTIONAL DESCRIPTION FRONT PANEL Keypad SET MEAS To display setpoint To display measurement OPERATIONAL GUIDE POWER CONNECTION The "ALARM" LED will light up when the alarm contact is closed, to indicate a TDS value higher than selected setpoint. CALIBRATION HI983315 LAYOUT Front view of the panel-mounted unit Trimmers For calibration CAL To adjust setpoint SET LEDs SET MEAS ALARM ON when the LCD displays set value ON when the LCD displays measured value ON when the alarm contact is closed HI 983315-0 and HI 98331500 Connect a 2-wire power cable to the terminal strip while paying attention to the correct positive and negative polarities (12 VDC). HI 983315-1 Connect a 3-wire power cable to the terminal strip while paying attention to the correct earth, neutral and line contacts (115/230 VAC). ALARM CONTACT Use this contact (maximum 5A, 240VAC, 30VDC) for connection to an alarm or dosing system. Notes: · All external cables connected to the rear panel should end with wire lugs. · It is recommended to cover the unused terminals with insulating tape. OPERATING THE METER All operations can be controlled via front panel keys and trimmers. [. . . ] · If a more thorough cleaning is required, brush the metal pins with very fine sandpaper. · After cleaning, rinse the probe with tap water and recalibrate the meter. · When not in use, clean the probe before storing it. Panel-mounted unit assembling view HI 740146 ACCESSORIES HI 7634-00 HI 70033P HI 7033M HI 7033L HI 7061M HI 710005 HI 710006 HI 710012 HI 710013 HI 710014 HI 731326 EC/TDS probe 84 µS/cm Calibration solution, 20mL 84 µS/cm Calibration solution, 230mL 84 µS/cm Calibration solution, 460mL Electrode cleaning solution, 230 mL 12 VDC power adapter, US plug 12 VDC power adapter, European plug 12 VDC power adapter, Australian plug 12 VDC power adapter, South African plug 12 VDC power adapter, UK plug Calibration Screwdriver (20 pcs) Note: Note: the power input is internally protected by a 400 mA fuse 2. [. . . ]

DISCLAIMER TO DOWNLOAD THE USER GUIDE HANNA INSTRUMENTS HI 983315-0




Click on "Download the user Manual" at the end of this Contract if you accept its terms, the downloading of the manual HANNA INSTRUMENTS HI 983315-0 will begin.

 

Copyright © 2015 - manualRetreiver - All Rights Reserved.
Designated trademarks and brands are the property of their respective owners.