If this document matches the user guide, instructions manual or user manual, feature sets, schematics you are looking for, download it now. Diplodocs provides you a fast and easy access to the user manual DHI INTERCONNECTION KIT FOR MPG1 WITH RPM4 OR PG 7302 PN 402265 INSTRUCTION SHEET. We hope that this DHI INTERCONNECTION KIT FOR MPG1 WITH RPM4 OR PG 7302 PN 402265 INSTRUCTION SHEET user guide will be useful to you.
Manual abstract: user guide DHI INTERCONNECTION KIT FOR MPG1 WITH RPM4 OR PG 7302 PN 402265 INSTRUCTION SHEET
Detailed instructions for use are in the User's Guide.
[. . . ] Interconnection Kit For MPG1 with RPM4 or PG 7302
P/N 402265 Instruction Sheet
INTERCONNECTION KIT FOR MPG1 WITH RPM4 OR PG 7302, P/N 402265, INCLUDES:
DESCRIPTION QTY PART NO.
DH500 6 in. NPTM
1 1 2 1 1 2
100208 100209 123019 100291 100285 102819
Install nipples into three way tee: Install the 5 in. [. . . ] Allow two nipple threads to show between the collar and nipple tip.
a
Figure 1: Two nipples installed in 3-way tee.
Optional: Extend the front feet on RPM4: The RPM4's front feet can be extended so that the RPM4 is inclined for easier reading and display access. Wrap Teflon thread tape around the NPT adapter's threads three times, and install the adapter into the RPM4. Install the free end of the center nipple into RPM4 or PG 7302: The center, thin, 5 in. nipple should be inserted into the RPM4's or the PG 7302's test port. The nipple can be bent slightly to ensure that nipple enters straight into the test port. Do not tighten the gland in the RPM4 or PG 7302 yet, since it will be tightened when the system is primed.
RPM4 Install glands and collars onto the 6 in. PG 7302 Install glands and collars onto the 12 in nipple.
b
Install the 6 in. Install a DH500M to 1/8" NPT male adapter in the DUT if necessary: The 1/8 in. If you intend to use the NPT adapter, wrap Teflon thread tape around its threads three times and install it into the DUT. The DUT will be connected to the test line by the remaining thin 5 in. [. . . ] There are 3 loose fittings: one on the DUT, one on the RPM4 or PG 7302, and the other is the plug. Tighten each of the loose fittings when oil without bubbles begins to flow through it.
Document 560080
050815
DH Instruments, Inc.
4765 East Beautiful Lane Phoenix AZ 85044-5318 USA
Tel 602. 431. 9100 Fax 602. 431. 9559 dhi@dhinstruments. com www. dhinstruments. com [. . . ]
DISCLAIMER TO DOWNLOAD THE USER GUIDE DHI INTERCONNECTION KIT FOR MPG1 WITH RPM4 OR PG 7302 PN 402265 INSTRUCTION SHEET
Click on "Download the user Manual" at the end of this Contract if you accept its terms, the downloading of the manual DHI INTERCONNECTION KIT FOR MPG1 WITH RPM4 OR PG 7302 PN 402265 INSTRUCTION SHEET will begin.