User manual CITIZEN CH-302B
DON'T FORGET : ALWAYS READ THE USER GUIDE BEFORE BUYING !!!
If this document matches the user guide, instructions manual or user manual, feature sets, schematics you are looking for, download it now. Diplodocs provides you a fast and easy access to the user manual CITIZEN CH-302B. We hope that this CITIZEN CH-302B user guide will be useful to you.
Manual abstract: user guide CITIZEN CH-302B
Detailed instructions for use are in the User's Guide.
[. . . ] 400 times/6F22 (once a day, room temperature 22°C) 9V DC +10°C~+40°C, 30%~85%RH 20°C~+60°C, 10%~95%RH 85mm(W) × 120mm(D) × 35mm(H) Approx. 150g (Apparatus and battery) Type B equipment Cuff, Battery (6F22), Instruction manual
*Specifications may be changed without notice in the event of improvements being made.
- Eng 2 -
English
CLEANING METHOD OF CUFF
After cleaning the cuff with a neutral detergent, lightly rinse it with water and let it dry in the air. Also, be careful not to get water into the air hose.
GENERAL REMARKS
General Remarks on blood pressure measurement
· If the cuff is to be deflated quickly, depress "POWER" switch. [. . . ] Do NOT measure blood pressure when it is too cold (below 10°C) or too hot (above 40°C) in the room. · Do NOT measure blood pressure immediately after drinking coffee or tea or smoking.
- Eng 3 -
· Measure blood pressure with a relaxed and quiet posture. Keep the center of the cuff at the level of your heart, and do NOT move your arm or talk during measurement.
· Do NOT keep on measuring so many times, as your arm will have built up congestion of blood, and you will not get a valid reading.
PRECAUTION FOR USE AND MAINTENANCE
Precautions for use
· If you suffer from heart disease, high blood pressure or other circulatory disease, please consult your doctor first. · If you feel abnormal pressure of the cuff during use or any other irregularity, reduce the pressure by immediately switching off the power or removing the batteries and consult the sales outlet where purchased. · If you think the measured value is abnormal or if measuring makes you feel unwell, discontinue use and consult your doctor. · Measurement of blood pressure may not be possible for someone with a weak pulse or arrhythmia. · If blood pressure is measured repeatedly, an accurate result will not be achieved. · The memory is erased if the batteries are removed for replacement, etc. and a record should be kept of any necessary data.
Precautions for maintenance
· Do NOT store the blood pressure monitor where it will be exposed to direct sunlight, high temperature (over 60°C), low temperature (below 20°C), high relative humidity (over 95%) or in a dusty place. · Do NOT expose the blood pressure monitor to under force or vibration, or drop it. · Remove the battery/batteries when not in use for a long period. · NEVER wipe the blood pressure monitor with thinner or benzene, as they may damage it. · When the monitor is very dirty, wipe it clean with a cloth moistened with Sterilizing alcohol or a neutral detergent, and then wipe it with a dry cloth. Do NOT use thinner or benzene.
Nature and frequency of maintenance:
This products is designed for use over an extended period of time; however, it is generally recommended that it be inspected every two years to ensure proper function and performance.
- Eng 4 -
Protect the nature environment:
Please help protect natural environment by respecting national and/or local recycling regulations when disposing of the batteries and the products at the end of their useful lives.
Manufacturer:
Name: Address: Premier Precision Ltd. Block E, G-H, 6th floor, Phase 2, Yip Fat Factory Building, 73-75 Hoi Yuen Road, Kwun Tong, Kowloon, Hong Kong
English
Reference European standard:
This products complies with EC instructions 93/42/EEC (Medical Device Directive) and with the following standards: EN 10601 : 1995 "Noninvasive Sphygmomanometers, Part 1 General Requirements" and EN 1060-3 : 1997 "Noninvasive Sphygmomanometers, Part 3 Supplementary Requirements" for Electromechanical Blood-Pressure Measuring Systems.
European Representative:
Name: Address: WvW elektronische Geräte Vertrieb GmbH. Robert-Koch-Straße 20, D-80538 München, GERMANY
PREPARATION FOR MEASUREMENTS
1. Inserting hoses
Bulb hose Cuff hose
Take off the battery cover at the back of the unit by sliding it as shown. Insert the cuff hose and bulb hose into the hose connector.
- Eng 5 -
3. [. . . ] Power off
SYSTOLIC mmHg DIASTOLIC mmHg
PULSE/min POWER
Flashing
SYSTOLIC mmHg DIASTOLIC mmHg
Touch the power switch for power off. If this unit is left on after measurement an automatic power-off function turns the power off after approximately 3 minutes.
PULSE/min
When the measurement is completed, the buzzer beeps, Systolic pressure is displayed on the left and Diastolic pressure on the right. Altemate display Approximately 2 seconds later, blood pressure values and pulse appear alternately on the display
Touch the power switch
Flashing
SYSTOLIC mmHg DIASTOLIC mmHg
PULSE/min
- Eng 8 -
English
DESCRIPTION OF DISPLAY MARKS
DISPLAY MARK
SYSTOLIC mmHg DIASTOLIC mmHg
CONDITION/CAUSE The cuff was not inflated to more than 30-40 mmHg above your systolic pressure.
CORRECTIVE ACTION Raise the set pressure value by higher level
PULSE/min SYSTOLIC mmHg DIASTOLIC mmHg
The cuff was inflated to more than 320 mmHg.
PULSE/min SYSTOLIC mmHg DIASTOLIC mmHg
Exhaust the air within the cuff.
Appears when the blood pressure could not be measured accurately.
PULSE/min SYSTOLIC mmHg DIASTOLIC mmHg
Repeat measurement again. [. . . ]